New extravascular defibrillator found safe, effective in global study: NEJM

Written By :  Dr. Kamal Kant Kohli
Published On 2022-09-06 05:30 GMT   |   Update On 2022-09-06 09:07 GMT

According to a global clinical study a new extravascular defibrillator using a lead placed behind the sternum has been found safe and effective.The device effectively terminated acute and chronic life-threatening ventricular arrhythmias.The findings were presented during a late-breaking session at the European Society of Cardiology Congress and were simultaneously published in The New...

Login or Register to read the full article

According to a global clinical study a new extravascular defibrillator using a lead placed behind the sternum has been found safe and effective.The device effectively terminated acute and chronic life-threatening ventricular arrhythmias.

The findings were presented during a late-breaking session at the European Society of Cardiology Congress and were simultaneously published in The New England Journal of Medicine.

Arrhythmias such as ventricular fibrillation and ventricular tachycardia happen in the lower heart chambers, or ventricles. They are dangerous because they interfere with the normal coordinated filling and pumping of blood through the heart. These arrhythmias can cause collapse and death if not rapidly treated. The ability of ICDs to accurately detect and terminate ventricular arrhythmias in high-risk patients saves lives.

For study participants, the lead of the extravascular ICD was inserted under the sternum, compared to transvenous ICD leads that are inserted through the veins into the heart, or subcutaneous ICDs that have a lead placed beneath the skin above the sternum. Patients with prior open-heart surgery, or who required pacing for a slow heart rate or had a pacemaker, were not candidates for this study.

This study is encouraging. By placing the lead in this new position behind the sternum, the uncommon but serious risks associated with transvenous ICDs, such as lung collapse, damage to heart valves and heart perforation, can be avoided. Limitations of the subcutaneous ICD are also overcome.

Since the lead is behind the sternum and close to the heart, pacing can be delivered, and defibrillation requires less energy with a longer battery life than with the subcutaneous ICD, says Paul Friedman, M.D., a cardiac electrophysiologist, and principal global investigator and first author of the study. Dr. Friedman chairs the Department of Cardiology at Mayo Clinic in Rochester.

The new extravascular ICD delivered anti-tachycardia pacing-rapid pacing-to painlessly terminate 70% of ventricular tachycardia episodes for which it was applied, (which is) at least as good as transvenous ICDs and not available in the subcutaneous ICD," says Dr. Friedman. "The extravascular ICD was also able to provide backup pacing to prevent pauses and was able to effectively defibrillate using a device half the size of the subcutaneous ICD."

Researchers from 17 countries participated in the study. Of the 316 patient participants with an attempted implant, 299 were discharged with a working extravascular ICD system. The defibrillation success rate was 98.7%. At six months, 92.6% of participants had no major system or procedure-related complications.

"It is important to note that electrophysiologists do not routinely place device leads behind the sternum. However, we found that with a robust training program, we could safely do so in the electrophysiology lab.

Initial implantations at each site included a cardiologist and a cardiac surgeon. The average procedure time was 74 minutes, similar to that of the early subcutaneous ICD experience," says Yong-Mei Cha, M.D., director of the Implantable Device Lab at Mayo Clinic and site principal investigator of the study.

"While encouraging, these results reflect an early experience, and there is more to be learned with longer follow-up and greater usage," says Dr. Friedman. "The rates of inappropriate shock are higher than with current devices but similar or lower than the early experience with other types of defibrillators, and steps have already been taken to lower it. Additionally, this new device is not for everyone. Patients with previous open-heart surgery were excluded from the study and would not be candidates for this treatment at present."

The study was sponsored by Medtronic Inc., which also developed the device. The extravascular ICD is premarket and has not been approved by the Food and Drug Administration. But it can be available, pending review, through a continued access study.

Reference:

Paul Friedman, Francis Murgatroyd, Lucas V.A. Boersma, Jaimie Manlucu, David O'Donnell, Bradley P. Knight, Nicolas Clémenty, Christophe Leclercq, Anish Amin, Béla P. Merkely, Ulrika M. Birgersdotter-Green, Joseph Y.S. Chan, et al. August 28, 2022 DOI: 10.1056/NEJMoa2206485

Tags:    
Article Source : New England Journal of Medicine,NEJM

Disclaimer: This site is primarily intended for healthcare professionals. Any content/information on this website does not replace the advice of medical and/or health professionals and should not be construed as medical/diagnostic advice/endorsement/treatment or prescription. Use of this site is subject to our terms of use, privacy policy, advertisement policy. © 2024 Minerva Medical Treatment Pvt Ltd