Postapproval study results of Impella show no mortality concerns: FDA
USA: In its recently released update, the U.S. Food and Drug Administration (FDA) has said that final postapproval study (PAS) results for the Impella RP system now alleviate concerns about a potential mortality signal previously seen with the percutaneous device. Also, the FDA updated the labelling for Abiomed's Impella RP System.
The final results from the PAS revealed that the survival rate for the subgroup of PAS patients who would have met the enrollment criteria for the premarket clinical studies is consonant with the premarket clinical studies survival rate. The findings further confirmed that the device is safe and effective when used for the currently approved indication.
Based on the PAS results, the FDA approved updated labelling for the Impella RP System on December 5, 2022, with an update of the indications for use statement to better consider the characteristics of the patients who may benefit the most from the device's treatment.
The results showed a survival rate of 69.7% for the PAS patients who would have been eligible for enrollment in premarket clinical studies. Survival was defined as living to 30 days hospital discharge or post explant or to receipt of a surgically implanted right ventricular assist device or heart transplant.
However, the survival rate for those who would not have met enrollment criteria was 18.6%. In these patients, the PAS survival rate should be interpreted in the context of conditions and limited treatment options. Such patients, according to the FDA, were more likely to have been in acute right heart failure or decompensation for longer than 48 hours and experienced a severe cardiogenic shock, end-organ failure, or acute neurologic injury.
"The Impella RP System is indicated for providing temporary right ventricular support for up to 14 days in patients with a body surface area ≥1.5 m2, who develop acute right heart failure or decompensation for less than 48 hours following left ventricular assist device implantation, myocardial infarction, heart transplant, or open‐heart surgery, without the presence of profound shock, end organ failure, or acute neurologic injury," the FDA wrote in its statement.
The FDA believes that when the device is used for the currently approved indication in appropriately selected patients the benefits of the Impella RP System continue to outweigh the risks.
The FDA encourages health care providers to give information on any adverse events or suspected adverse events associated with the use of Impella RP System. "Prompt reporting of adverse events can help the FDA identify and better understand the risks associated with medical devices," the article stated.
Disclaimer: This website is primarily for healthcare professionals. The content here does not replace medical advice and should not be used as medical, diagnostic, endorsement, treatment, or prescription advice. Medical science evolves rapidly, and we strive to keep our information current. If you find any discrepancies, please contact us at corrections@medicaldialogues.in. Read our Correction Policy here. Nothing here should be used as a substitute for medical advice, diagnosis, or treatment. We do not endorse any healthcare advice that contradicts a physician's guidance. Use of this site is subject to our Terms of Use, Privacy Policy, and Advertisement Policy. For more details, read our Full Disclaimer here.