Methacrylate Threadlockers found best for Enhancing Abutment Screw Retention: Study

Written By :  Dr. Shravani Dali
Medically Reviewed By :  Dr. Kamal Kant Kohli
Published On 2026-01-08 15:00 GMT   |   Update On 2026-01-09 07:23 GMT
Advertisement

Researchers have found in a new study that threadlockers significantly improved abutment screw retention, with methacrylate outperforming dimethacrylate. Reduced detorque under cyclic loading emphasizes the need for improved retention strategies, and methacrylate-based threadlockers may help prevent screw loosening and improve long-term implant success.

A study was done to evaluate and compare the effects of different thread lockers on implant abutment screw loosening. Thirty implant-abutment assemblies were divided into 3 groups: Control group (without threadlocker), dimethacrylate group, and methacrylate group. Each group was subdivided into static and dynamic groups. The specimens were torqued to 35 Ncm, and detorque values were measured after 1 month in static conditions and after cyclic loading. The cyclic loading protocol involved 240,000 cycles at a velocity of 16 mm/s over a 48-h period in a moist environment, simulating 1 year of clinical function.

Advertisement

Results were assessed using one-way, two-way ANOVA and multiple pairwise comparisons with a significance level of P ≤ 0.05. The methacrylate group demonstrated the highest detorque values (static: 36.46 ± 0.95 Ncm, dynamic: 33.22 ± 0.34 Ncm), followed by the dimethacrylate group (static: 32.84 ± 1.12 Ncm, dynamic: 29.98 ± 1.05 Ncm). The control group exhibited the lowest detorque values (static: 28.80 ± 0.96 Ncm, dynamic: 25.84 ± 0.63 Ncm). Detorque values were consistently higher in the static condition (32.40 ± 3.37 Ncm) compared to the dynamic condition across all groups (29.68 ± 3.20 Ncm), with statistically significant differences observed (P = 0.000).

Within the limitations of the study, application of threadlockers significantly improved abutment screw retention, with methacrylate exhibiting superior performance over dimethacrylate. The reduction in detorque values under cyclic loading highlights the need for enhanced retention strategies in implant prosthetics. Incorporating methacrylate-based threadlockers in clinical practice may mitigate screw loosening and enhance the long-term success of implant-supported prostheses.

Reference:

Arul, Lizy Teena, et al. "Comparative Evaluation of Abutment Screw Loosening With Two Different Thread Lockers Under Cyclic Loading: an in Vitro Study." Journal of Indian Prosthodontic Society, vol. 26, no. 1, 2026, pp. 49-55.

Tags:    
Article Source : Journal of Indian Prosthodontic Society

Disclaimer: This website is primarily for healthcare professionals. The content here does not replace medical advice and should not be used as medical, diagnostic, endorsement, treatment, or prescription advice. Medical science evolves rapidly, and we strive to keep our information current. If you find any discrepancies, please contact us at corrections@medicaldialogues.in. Read our Correction Policy here. Nothing here should be used as a substitute for medical advice, diagnosis, or treatment. We do not endorse any healthcare advice that contradicts a physician's guidance. Use of this site is subject to our Terms of Use, Privacy Policy, and Advertisement Policy. For more details, read our Full Disclaimer here.

NOTE: Join us in combating medical misinformation. If you encounter a questionable health, medical, or medical education claim, email us at factcheck@medicaldialogues.in for evaluation.

Our comments section is governed by our Comments Policy . By posting comments at Medical Dialogues you automatically agree with our Comments Policy , Terms And Conditions and Privacy Policy .

Similar News