Novel electronic pulp sensitivity tester with transilluminator efficient and safe method to heat test pulp
A novel electronic pulp sensitivity tester with a transilluminator efficient and safe method to heat test pulp suggests a new study published in the Journal of Endodontics.
The efficacy and safety of a single novel electronic pulp sensitivity tester with a transilluminator (PSTT) capable of providing 4 different tests were compared with gold standard (GS) pulp testing methods.
Four hundred eighty teeth, including 3 from each quadrant and their contralateral and opposing teeth, were randomly assessed using the PSTT or GS methods. Seven days later, the same teeth were assessed using the method not used earlier so that all teeth were evaluated using both methods. Sixty previously root canal–treated teeth, serving as negative controls, were assessed identically to the experimental groups. Results were analyzed using IBM SPSS software (IBM Corp, Armonk, NY) (P < .05).
Results
A higher percentage of teeth responded to cold and electricity using GS methods compared with the PSTT (99.4 vs 93.1 and 99.6 vs 97.3, respectively). A lower percentage of teeth responded to heat using the GS method compared with the PSTT (50.0 vs 68.1). Chi-square tests determined differences in the proportion of sensitivity to temperature, and electric testing methods were statistically significant (P < .001 and P = .004, respectively). No difference in the proportion of sensitivity was observed for transillumination. The Wilcoxon signed rank test determined significantly shorter cold testing times using the GS method (P = .024). Shorter testing times were observed using the PSTT for heat (P < .001), electric pulp testing (P = .048), and transillumination (P = .001). The overall PSTT testing time was significantly shorter than the GS testing time (P = .03). Tissue injury was not observed.
The PSTT efficiently and safely provided heat and transillumination for pulpal diagnosis. Improvements are needed to enhance the cold and electric stimulus efficacy.
Reference:
Comparison of the Efficacy and Safety of Pulp Sensitivity Testing Results Using a New Pulp Testing Device versus Some Commonly Used Methods: A Crossover Study
Patricia A. Tordik, Mahmoud Torabinejad, Cassandra M. Fevelo, Mason Bahador, DDS, MS, Udochukwu Oyoyo, Published:October 13, 2023DOI:https://doi.org/10.1016/j.joen.2023.10.003
Keywords:
Novel, electronic, pulp, sensitivity, tester, transilluminator, efficient, safe, method, heat, test, pulp, Testing Device versus Some Commonly Used Methods: A Crossover Study
Patricia A. Tordik, Mahmoud Torabinejad, Cassandra M. Fevelo, Mason Bahador, DDS, MS, Udochukwu Oyoyo
Disclaimer: This website is primarily for healthcare professionals. The content here does not replace medical advice and should not be used as medical, diagnostic, endorsement, treatment, or prescription advice. Medical science evolves rapidly, and we strive to keep our information current. If you find any discrepancies, please contact us at corrections@medicaldialogues.in. Read our Correction Policy here. Nothing here should be used as a substitute for medical advice, diagnosis, or treatment. We do not endorse any healthcare advice that contradicts a physician's guidance. Use of this site is subject to our Terms of Use, Privacy Policy, and Advertisement Policy. For more details, read our Full Disclaimer here.