BLU DAT: new reliable and accurate arthrometer for measuring anterior knee laxity

Published On 2023-12-31 22:00 GMT   |   Update On 2024-01-01 07:18 GMT
Advertisement

Although many arthrometers have been developed to assess anterior knee laxity, reliability and diagnostic accuracy of these devices are still debated.

Alessandro Colosio et al conducted a study to evaluate the validity of a new arthrometer in the outpatient setting, with the hypothesis that it had good validity in terms of reliability and diagnostic accuracy.

The study was conducted at Department of Bone and Joint Surgery, Spedali Civili, Brescia, Italy. It has been published in “Knee Surgery, Sports Traumatology, Arthroscopy journal.”

Advertisement

Seventy-eight subjects (39 with ACL injury and 39 with normal ACL) were tested. ATT was assessed by means of the Lachman test at 30° of flexion with a new testing device (BLU-DAT) under three different loading conditions: 7 kg (69 N), 9 kg (88 N) and maximum (MMT). The tests were performed on both knees to obtain SSD. In the ACL injury group, the tests were performed by two examiners and one of them repeated a second test series. Inter- and intra-observer reliability were assessed with the intraclass correlation coefficients (ICCs) for the average SSD measures. In the normal-ACL group, the analysis was performed with the same testing setup. Side-to-side difference measures of the two groups at every loading condition were compared by Student’s t test. Data of test series were dichotomized based on the threshold value of 3-mm SSD as pathological ATT and 2×2 contingency tables were used to assess diagnostic accuracy.

Key findings of the study were:

The ICCs for intra-observer reliability at 7-kg (69 N), 9-kg (88 N) and MMT measurements were 0.781, 0.855 and 0.913, respectively.

The ICC for inter-observer reliability at 7-kg (69 N), 9-kg (88 N) and MMT measurements were 0.701, 0.845 and 0.834, respectively.

Comparison between the two groups showed a significant mean difference ranging from 3.4 mm for 7-kg (69 N) load to 4.6 mm for MMT.

Overall accuracy ranged from 84.6% for 7-kg load to 98.7% for MMT.

The authors concluded that – “The BLU-DAT has proven to be an instrument with good intra- and inter-observer reliability and very good accuracy in the diagnosis of ACL injuries in the outpatient setting. So, the BLU-DAT can be a new useful tool in everyday clinical practice to assist in the diagnosis of ACL injury.”

Level of evidence II.

Further reading:

BLU DAT: a new reliable and accurate arthrometer for measuring anterior knee laxity

Alessandro Colosio, Alessandra Scaini et al

Knee Surgery, Sports Traumatology, Arthroscopy (2023) 31:4782–4790

https://doi.org/10.1007/s00167-023-07534-5

Tags:    
Article Source : “Knee Surgery, Sports Traumatology, Arthroscopy journal.”

Disclaimer: This website is primarily for healthcare professionals. The content here does not replace medical advice and should not be used as medical, diagnostic, endorsement, treatment, or prescription advice. Medical science evolves rapidly, and we strive to keep our information current. If you find any discrepancies, please contact us at corrections@medicaldialogues.in. Read our Correction Policy here. Nothing here should be used as a substitute for medical advice, diagnosis, or treatment. We do not endorse any healthcare advice that contradicts a physician's guidance. Use of this site is subject to our Terms of Use, Privacy Policy, and Advertisement Policy. For more details, read our Full Disclaimer here.