Self-compression mammography gives same image quality as standard compression: Study
Italy: Mammograms performed with self-compression compared to radiographer-led compression of breasts helped to achieve stronger breast compression with comparable image quality, says a recent study published in European Radiology. Also, the proportion of women who attended follow-up mammography eas also found to be similar in the two groups.
Appropriate compression is said to be essential in mammography in order to obtain high image quality and reduce dose. compression however can cause pain and discomfort. It has been proposed that self-compression reaches better compression and possibly increase participation in mammography.
Against the above background, Paolo Giorgi Rossi, Azienda Unità Sanitaria Locale – IRCCS di Reggio Emilia, via Amendola Reggio Emilia, Italy, and colleagues conducted the study with the objective to test the efficacy of self- compared to radiographer-led compression to reduce the average glandular dose without affecting image quality and compliance to follow-up mammography.
For this purpose, the researchers included women presenting for mammography for breast cancer follow-up, symptoms, opportunistic screening, or familial risk. They were randomized to self-compression or radiographer-led compression. Two independent radiologists and two radiographers blindly assessed image quality. Measurement of pain and discomfort was done immediately after mammography and their recall was asked when the women participated in the follow-up mammogram, 1 or 2 years later.
Based on the study, the researchers reported the following findings:
- In total, 495 women (mean age 57 years +/-14) were enrolled, 245 in the self-compression and 250 radiographer-compression arms.
- Image quality was similar in the two arms.
- A stronger compression force was reached in the self- than in the radiographer-arm (114.5 vs. 10.25 daN), with a 1.7-mm reduction in thickness, and almost no impact on dose per exam (1.90 vs. 1.93 mGy).
- Moderate/severe discomfort was reported by 7.8% vs 9.6% and median pain score was 4.0 in both arms.
- Median execution time was 1 min longer with self-compression (10.0 vs. 9.1 min).
- No effect on subsequent mammography was detectable.
To conclude, self-compression reached stronger compression of the breast, with comparable image quality but with no glandular dose reduction or impact on participation in follow-up mammography."
Reference:
Iotti, V., Giorgi Rossi, P., Canovi, L. et al. Patient-centred care with self-compression mammography in clinical practice: a randomized trial compared to standard compression. Eur Radiol (2022). https://doi.org/10.1007/s00330-022-09002-z
Disclaimer: This website is primarily for healthcare professionals. The content here does not replace medical advice and should not be used as medical, diagnostic, endorsement, treatment, or prescription advice. Medical science evolves rapidly, and we strive to keep our information current. If you find any discrepancies, please contact us at corrections@medicaldialogues.in. Read our Correction Policy here. Nothing here should be used as a substitute for medical advice, diagnosis, or treatment. We do not endorse any healthcare advice that contradicts a physician's guidance. Use of this site is subject to our Terms of Use, Privacy Policy, and Advertisement Policy. For more details, read our Full Disclaimer here.