3D printed patient specific instrumentation accurate tool to reproduce femoral bone tunnels in multiple ligament knee injuries
Multiple-ligament knee reconstruction techniques often involve the creation of several bone tunnels for various reconstruction grafts. A critical step in this procedure is to avoid short tunnels or convergences among them. Currently, no specific template guide to reproduce these angulations has been reported in the literature, and the success of the technique still depends on the experience of the surgeon.
Núria Fernández Poch et al conducted a study to analyze the accuracy and reliability of 3D-printed patient-specific instrumentation (PSI) for lateral and medial anatomical knee reconstructions.
10 cadaveric knees were scanned by computed tomography (CT). Using specific computer software, anatomical femoral attachments were identified:
(1) On the lateral side the lateral collateral ligament (LCL) and the popliteal tendon (PT) and
(2) On the medial side the medial collateral ligament (MCL) and the posterior oblique ligament (POL).
Four bone tunnels were planned for each knee, and PSI with different directions were designed as templates to reproduce the planned tunnels during surgery. 20 3D-printed PSI were used: ten were tailored to the medial side for reconstructing MCL and POL tunnels, and the other ten were tailored to the lateral side for reconstructing LCL and PT tunnels. Postoperative CT scans were made for each cadaveric knee. The accuracy of the use of 3D-printed PSI was assessed by superimposing postoperative CT images onto pre-operative images and analyzing the deviation of tunnels performed based on the planning, specifically the entry point and the angular deviations.
Key findings of the study:
• The median entry point deviations for the tunnels were as follows:
LCL tunnel, 1.88 mm (interquartile range (IQR) 2.2 mm);
PT tunnel, 2.93 mm (IQR 1.17 mm);
MCL tunnel, 1.93 mm (IQR 4.26 mm); and
POL tunnel, 2.16 mm (IQR 2.39) .
• The median angular deviations for the tunnels were as follows:
LCL tunnel, 2.42° (IQR 6.49°);
PT tunnel, 4.15° (IQR 6.68);
MCL tunnel, 4.50° (IQR 6.34°); and
POL tunnel, 4.69° (IQR 3.1°) .
• No statistically significant differences were found in either the entry point or the angular deviation among the different bone tunnels.
The authors concluded that – “The use of 3D-printed PSI for lateral and medial anatomical knee reconstructions provides accurate and reproducible results and may be a promising tool for use in clinical practice.”
Further reading:
Three dimensional printed patient specific instrumentation is an accurate tool to reproduce femoral bone tunnels in multiple ligament knee injuries
Núria Fernández Poch, Ferran Fillat Gomà et al
International Orthopaedics (2023) 47:1213–1219
https://doi.org/10.1007/s00264-023-05712-1
Disclaimer: This website is primarily for healthcare professionals. The content here does not replace medical advice and should not be used as medical, diagnostic, endorsement, treatment, or prescription advice. Medical science evolves rapidly, and we strive to keep our information current. If you find any discrepancies, please contact us at corrections@medicaldialogues.in. Read our Correction Policy here. Nothing here should be used as a substitute for medical advice, diagnosis, or treatment. We do not endorse any healthcare advice that contradicts a physician's guidance. Use of this site is subject to our Terms of Use, Privacy Policy, and Advertisement Policy. For more details, read our Full Disclaimer here.