Use of Digital Fluroscopy resulted in more accurate execution of lengthening in THA
Limb length discrepancy (LLD) after THA is an undesirable outcome, and its incidence in the literature varies between 1% and 27 %. It can lead to patient dissatisfaction and secondary complications including gait disturbance, low-back pain, neuritis, component instability, and early component loosening. Clinical evaluation and preoperative templating establish the intended lengthening.
R O’Leary et al conducted a study to assess whether digital fluoroscopic navigation (DF) improved the accuracy of planned lengthening in direct anterior approach total hip arthroplasty (DAA-THA).
The study has been published in ‘Arthroplasty Today’ journal.
Planned lengthening measurements on 100 consecutive unilateral DAA-THA patients, along with patient characteristics, were prospectively collected by 2 surgeons. One surgeon utilized DF to achieve intended length (n = 50), while the other utilized unaided standard fluoroscopy (SF; n = 50). A third surgeon blinded to the procedures assessed actual limb length using an ipsilateral overlay technique on the 6-week postoperative radiograph. The difference between the mean planned and actual limb lengthening stratified by DF and SF was assessed using bivariate and multivariate statistics.
Key findings of the study were:
• The mean (SD) age was 65.0 (9.7) years (range, 46-87).
• Osteoarthritis was the diagnosis in 95%, while avascular necrosis was seen in 5%.
• Both groups had similar demographic characteristics.
• 67 procedures were right-sided, and 33 procedures were of left laterality.
• The mean (standard deviation) planned lengthening in DF and (standard unaided fluoroscopy) SF groups was 3.96 (2.1) and 3.47 (2.2) mm, respectively.
• The mean (standard deviation) actual lengthening in DF and SF groups was 3.11 (4.0) and 0.68 (4.6) mm, respectively.
• The mean (SD) BMI for patients in DF and SD groups was 29.72 (6.5) and 29.39 (11.8).
• After accounting for age, sex, body mass index, laterality, and the Bone Index, multivariate regression results showed that the average difference between planned and actual limb lengthening in the DF group was significantly lower than that in the SF group (beta = 1.92; 95% confidence interval: 3.51, 0.33; P < .02).
• A greater percentage of patients in the DF group (66% vs 40%) were within 3 mm of the intended plan (P < .01).
The authors concluded that – “The use of fluoroscopy in DAA-THA helps achieve intended limb lengthening. Enhancement with digital navigation resulted in more accurate execution of planned limb lengthening.”
Further reading:
Digital Fluoroscopic Navigation for Limb Length Restoration During Anterior Total Hip Arthroplasty Robert O’Leary, Anshul Saxena et al Arthroplasty Today 18 (2022) 11-15 https://doi.org/10.1016/j.artd.2022.08.021
Disclaimer: This website is primarily for healthcare professionals. The content here does not replace medical advice and should not be used as medical, diagnostic, endorsement, treatment, or prescription advice. Medical science evolves rapidly, and we strive to keep our information current. If you find any discrepancies, please contact us at corrections@medicaldialogues.in. Read our Correction Policy here. Nothing here should be used as a substitute for medical advice, diagnosis, or treatment. We do not endorse any healthcare advice that contradicts a physician's guidance. Use of this site is subject to our Terms of Use, Privacy Policy, and Advertisement Policy. For more details, read our Full Disclaimer here.